11
STATus Subsystem
STATus:OPERation
The operation status register set contains conditions which are a part of
the operation of the power meter as a whole.
The following bits in these registers are used by the power meter:
Syntax
476
Bit Number
Decimal
Weight
0
1
1 - 3
-
4
16
5
32
6 - 9
-
10
1024
11
2048
12
4096
13 to 15
-
:OPER
STAT
Definition
CALibrating Summary
Not used
MEASuring Summary
Waiting for TRIGger Summary
Not used
SENSe Summary
Lower Limit Fail Summary
Upper Limit Fail Summary
Not used (bit 15 always 0)