Pstatistic:ccdf:reference:power:ptaverage - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

PSTatistic:CCDF:REFerence:POWer:PTAVerage?

This command is used to retrieve peak to average data of the saved
reference trace.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
PST:CCDF:REF:POW:PTAV?
Error Messages
• If command is executed in other window besides CCDF window, error
• If there was no previously saved trace, error –221 "Settings conflict: No
N1911A/1912A P-Series Power Meters Programming Guide
PST
:CCDF
–221 "Settings conflict: Requires CCDF window" occurs.
reference trace saved" occurs. Please check the status of saved reference
trace using command PST:CCDF:STOR:REF?.
:POW
:REF
This command returns the peak to
average power of the saved reference
trace.
PSTatistic Subsystem
:PTAV
?
9
329