Parameters
Example
TRAC:DATA? HRES
Error Messages
If TRAC:STAT is off, the error –221, "Settings Conflict" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
Item
Description/Default
character_data
HRESolution: high resolution. The complete capture
buffer at the internal sample rate. The number of points in
this trace is not fixed, as it is affected by the
SENS:TRACe:TIMe setting.
MRESolution: medium resolution. A subset of the
capture buffer - the buffer contents are decimated
data points.
LRESolution: low resolution. A subset of the capture
buffer - the buffer contents are decimated
data points. This is the same number of data points as the
power meter uses to display the trace on the front panel.
Hence, the LRES command can be used to replicate the
power meter's display.
1
to provide 230
This command returns the trace data for
Channel A at high resolution.
TRACe Subsystem
13
Range of
Values
HRES
MRES
LRES
1
to 1000
613