Example
BUFF:MTYP AVER
Query
[SENSe[1]]|SENSe2:BUFFer:MTYP?
This query is used to retrieve measurement type settings.
Query Example
BUFF:MTYP?
On Reset
On *RST, the value is set to AVER
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241,
• If E9320 or N1920 sensor is connected but acquisition mode is in free
• If E9320 or N1920 sensor is connected but sensor mode is average,
• If E9320 or N1920 sensor is connected but trigger source is not
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
run, error –221, "Setting conflict. Invalid acquisition mode" occurs.
error –221, "Setting conflict. Invalid acquisition mode" occurs.
external, error- 221 "Setting conflict. Invalid acquisition mode" occurs.
This command sets the measurement type
to average for Channel A.
This query returns the measurement type
to average for Channel A.