Example
BUFF:MTYP AVER
Query
[SENSe[1]]|SENSe2:BUFFer:MTYP?
This query is used to retrieve measurement type settings.
Query Example
BUFF:MTYP?
On Reset
On *RST, the value is set to AVER
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241,
• If E9320 or N1920 sensor is connected but acquisition mode is in free
• If E9320 or N1920 sensor is connected but sensor mode is average,
• If E9320 or N1920 sensor is connected but trigger source is not
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
run, error –221, "Setting conflict. Invalid acquisition mode" occurs.
error –221, "Setting conflict. Invalid acquisition mode" occurs.
external, error- 221 "Setting conflict. Invalid acquisition mode" occurs.
This command sets the measurement type
to average for Channel A.
This query returns the measurement type
to average for Channel A.
SENSe Subsystem
10
385