SERVice:SENSor[1]|2:POWer:PEAK:MAXimum?

This query returns the maximum peak power that can be measured by the
currently connected sensor. It is applicable to E- Series sensors only.
Maximum peak power information is stored in the sensor's EEPROM.
Syntax
Example
SERV:SENS2:POW:PEAK:MAX? This query returns the maximum peak
Error Messages
• If no sensor is connected, error –241, "Hardware missing" occurs.
• If a a sensor other than an E- Series sensor is connected, error –241
• If the E- Series sensor currently connected does not contain the
N1911A/1912A P-Series Power Meters Programming Guide
SERV
:SENS
"Hardware missing" occurs.
necessary information in EEPROM, error –241 "Hardware missing"
occurs.