STATus Subsystem

STATus:OPERation:LLFail[:SUMMary]

The operation status lower limit fail summary register set contains
information on the lower limit fail status of the power meter.
The following bits in these registers are used by the power meter:
The appropriate bits are set if a channel lower limit test fails or a window
lower limit test fails.
These bits are cleared if a measurement is made and the test is enabled
and passes.
Syntax
478
Bit
Decimal
Number
Weight
0
-
1
2
2
4
3
8
4
16
5
32
6
64
7-15
-
STAT
:OPER
Definition
Not used
Channel A LLFail Status
Channel B LLFail Status (N1912A only)
Upper window LLFail Status
Lower widow LLFail Status
Upper window lower measurement LLFail Status
Lower window lower measurement LLFail Status
Not used
:LLF
:SUMM