Trigger[:Sequence]:Hysteresis <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

TRIGger[:SEQuence]:HYSTeresis <numeric_value>
This command sets:
• How far a signal must fall below TRIG:LEVel before a rising edge can
• How far a signal must rise above TRIG:LEVel before a falling edge can
Syntax
Parameters
N1911A/1912A P-Series Power Meters Programming Guide
be detected.
be detected.
TRIG
:SEQ
Item
Description/Default
numeric_value
How far a signal must fall/rise before a
rising or falling edge can be detected.
Units are resolved to 0.05 dB.
:HYST
Space
DEF: the default value is 0 dB
TRIGger Subsystem
numeric_value
DEF
?
Range of Values
0 to 3 dB
DEF
14
669

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents