14
TRIGger Subsystem
Example
TRIG:SEQ:LEV 10
Reset Condition
On reset the power level is set to 0 dBm.
Query
TRIGger[:SEQuence]:LEVel?
The query returns the current power level setting.
Query Example
TRIG:SEQ1:LEV?
672
This command sets the power level for a
trigger event to 10 dBm.
This command queries the power level
setting for Channel A.
N1911A/1912A P-Series Power Meters Programming Guide