Query Example
CORR:GAIN2?
Error Messages
• If a loss/gain correction value is entered using
• The SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, "Settings
Conflict" occurs. However, the correction value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not
automatically set ON.
command can be used for the 8480 Series power sensor when no
sensor calibration table has been set up.
This command queries the current setting
of the channel offset on Channel A.