Trigger[:Sequence[1]|2]:Immediate - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

14
TRIGger Subsystem

TRIGger[:SEQuence[1]|2]:IMMediate

This command provides a one time over- ride of the normal process of the
downward path through the wait for trigger state. It causes the immediate
exit of the event detection layer if the trigger system is in this layer when
the command is received. In other words, the instrument stops waiting for
a trigger and takes a measurement ignoring any delay set by TRIG:DELay.
Syntax
Example
TRIG:SEQ:IMM
682
TRIG
:SEQ
1
:IMM
2
This command initiates a measurement
on Channel A.
N1911A/1912A P-Series Power Meters Programming Guide