10
SENSe Subsystem
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO:REF1|REF2
<numeric_value>
This command is used to set the Reference 1 and 2 of the selected gate
for Auto Gating Marker.
This command is only applicable when N192x or E932x sensor is present and trigger
N O T E
acquisition mode is selected.
Syntax
SENS
1
SENS2
Parameters
436
:
SWE
1
2
3
4
Item
Description/Default
numeric_value
The values of Auto Gating Marker References 1
and 2 for the selected gate.
The combined value of REF1 and REF2 can not
exceed 99.9 %.
:AUTO
:REF1
:REF2
N1911A/1912A P-Series Power Meters Programming Guide
Space
numeric_value
?
Range of Values
0.0 to 99.9