10
SENSe Subsystem
• 1 is returned when the pulse power measurement feature is enabled
• 0 is returned when the pulse power measurement feature is disabled
Query Example
CORR:GAIN3:STAT?
Error Messages
• If [SENSe[1]]|SENSe2:CORRection:DCYCle:STATus is set to ON while
• If this command is used when an E4412A/E4413A power sensor is
• If this command is used when a E9320 power sensor is set to NORMal
• If this command is used when a P- Series power sensor is connected,
402
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, "Settings
Conflict" occurs.
connected, the error –310, "System error;Dty Cyc may impair accuracy
with ECP sensor" occurs. If you are using a dual channel power meter
the error message specifies the channel.
mode, the error –221, "Settings Conflict" occurs.
the error –241, "Settings Conflict" occurs.
This command queries whether the pulse
measurement feature is on or off.
N1911A/1912A P-Series Power Meters Programming Guide