System:help:headers - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

SYSTem:HELP:HEADers?

This query returns a list of all SCPI commands supported by the
instrument.
Data is returned in IEEE 488.2 arbitrary block program data format as
shown in
Figure 12-21IEEE 488.2 Arbitrary Block Program Data Format
Each point in the trace is represented as an IEEE 754 32 bit floating
point number, made up of four bytes in the data block. The MS byte is
transmitted first. Each complete block is terminated by a line feed.
Commands are listed in alphabetical order.
Syntax
N1911A/1912A P-Series Power Meters Programming Guide
Figure 12- 21
#xyyy..yddd................ddd<LF>
The number of data bytes (d)
contained in the block.
The number of y digits
Signifies the start of the block
Example: if there are 12435 data bytes, y = 12435 and x = 5
:HELP
SYST
below.
?
:HEAD
SYSTem Subsystem
Line feed character
signifies the end of the block
Data bytes
12
523