Reset Condition
On reset, the CCDF cummulative counts will be set to the default value,
100 M samples.
Query
PSTatistic[1]|2:CCDF:COUNt?
The query returns the current numeric value of the CCDF cummulative
count for the respective channel selected.
Query Example
PST1:CCDF:COUN?
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241
• If channel is not in free- run acquisition mode, error –221 "Settings
• If user inputs a count that is lesser than 100 M or greater than 10 G,
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
conflict" occurs.
error –222 "Data out of range" occurs. Any input count that is within
range will be normalized and round down to the nearest 100 M.
This command queries the numeric value
of CCDF cummulative counts for Channel
A.