Power Meter Operation

Recorder Output

Recorder Output
The rear panel Recorder Output connector produces a dc voltage that corresponds to
the power level in Watts of the channel, depending on the measurement mode. This dc
voltage ranges from 0 to +1 Vdc. The output impedance is typically 1 kΩ . Channel
and display offsets, and duty cycle have no effect on the Recorder Output.
For example, the Recorder Output can be used to;
record swept measurements on an X-Y recorder
level an output from a source using external leveling, or
monitor the output power on a strip chart recorder. A setup for recording
swept measurements is shown in Figure 2-20.
See Figure 2-22 for details on which power meter functions are implemented in the
recorder output.
Figure 2-20: Test Setup for Recording Swept Measurements
Swept Source
SWEEP
RF OUT
OUT
X-AXIS (FREQUENCY)
To access the "Recorder" menu press
menu allows you to switch the Recorder Output signal either on or off. The
and
Max Power
Min Power
that you want to represent the 1 V
the Recorder Output.
2-58
Device
Under Test
IN
OUT
X-Y Recorder
System
,
More
Inputs
softkeys allow you to enter the input power level
maximum and 0 V
dc
RECORDER
OUTPUT
Power Meter
CHANNEL A
INPUT
Y-AXIS (POWER)
,
Recorder Output
minimum output voltage of
dc
Agilent E4418B User's Guide
. This