A3 Output AGC Gain Adjustment

The purpose of this procedure is to adjust the gain of AGC on the A3 Source Board.

Required Equipment

Calibration Kit
7mm-N(m) Adapter

Procedure

Step 1. Run the adjustment program.
Step 2. Choose the A3 Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Connect the instrument as shown in Figure 3-11.
Figure 3-11
A3 AGC Gain Adjustment Setup
Step 4. Follow the adjustment program instruction to update the correction constant.
Chapter 3
Description
16195B(furnished with E4991A)
11525A
A3 Output AGC Gain Adjustment
Recommended Model