2-8-2. POWER SUPPLY OUTPUT VOLTAGE ADJUSTMENT

This procedure adjusts the output voltage of the power supply.
EQUIPMENT:
Multimeter
2 Alligator Clips to 1 Dual Banana Plug
Test Lead
PROCEDURE:
1. Set the Multimeter to the DC voltage measurement function.
2. Connect the Test Leads to the multimeter.
3. Connect multimeter's L o input to A7TP2 (GND), and connect the Hi input to
A7TP1. Figure 2-7 shows the location of A7TP1 and A7TP2.
MULTIMETER
0
0
0
0
0
0
O f = 0 0
0
0
0
M C 1 1 = 3 0 0 0
I
I
Figure 2-6. Power Supply Voltage Adjustment Setup
A7TP1
A7TP2
(GND)
Figure 2-7. Power Supply Voltage Adjustment Location