GO/NO-GO Test with Limit Test Function

Use the limit test function to perform GO/NO-GO test. This function sets the upper and lower
limits to determine whether DUTs are acceptable. The limit di erentiation function compares
the measured data with preset limits and shows through its output and on-screen message
whether the DUT is within the limits.

Editing Limit Table

This section describes how to edit the limit table of primary parameters and secondary
parameters, as shown in Table 4-1.
Segment Stimulus Value
Testing function is not actually performed.
1
Note
If you do not want to perform testing function for upper or lower limits
of secondary parameters, specify suciently large value for upper limit or
suciently small value for lower limit.
1. Press
, and then
4
Test Setup
5
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT PRI LIMIT .
3. Segment 1
NNNNNNNNNNNNNN
Press
EDIT and
Enter Stimulus (1MHz).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
LOWER LIMIT .
Enter lower limit value (10nH).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT .
Press
Enter upper limit value (12nH).
NNNNNNNNNNNNNN
Press
DONE to complete the editing of this segment.
4. Segment 2
NNNNNNNNNNN
Press
ADD to edit the new segment.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
STIMULUS VALUE .
Enter Stimulus (10MHz).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
LOWER LIMIT .
Enter lower limit value (12nH).
4-4
4286A with Chip Handler
Table 4-1. Example of limit setting
Primary Parameter
Lower limit Upper limit Lower limit Upper limit
1
1MHz
2
10MHz
3
100MHz
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TEST MENU .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE .
Secondary Parameter
10nH
12nH
12nH
14nH
10nH
14nH
1
100.0
1G
1
100.0
1G
1
100.0
1G