OUTPut:RECorder[1]|2:LIMit:UPPer <numeric_value>
This command sets the maximum scaling value for the specified recorder
output. The units used are dependent on the units currently set for the
CALC block specified in OUTPut:RECorder[1]|2:FEED
<data_handle>.
Syntax
OUTP
:REC
Parameters
Item
numeric_value
Example
OUTP:REC:LIM:UPP 10
Reset Condition
On reset, the maximum scaling value is set to +20 dBm.
Query
OUTPut:RECorder[1]|2:LIMit:UPPer?
The query command returns the maximum scaling value.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OUTPut:RECorder[1]|2:LIMit:UPPer <numeric_value>
:LIM
:UPP
1
2
Description/Default
A numeric value for the maximum
scaling value. The units used—dBm, W
or %—are dependent on the units
currently set for the CALC block
specified in
OUTPut:RECorder[1]|2:FEED
<data_handle>.
OUTput Subsystem
numeric_value
Space
?
Range of Values
-150 to +20 dBm
1 aW to 100 mW
0% to 100%
This command sets the maximum
scaling value to 10.