Each string parameter returned indicates the name, type and size of a
stored frequency dependent offset table:
• <string>,<type>,<size>
The <string>, <type> and <size> are all character data. The <type>
is always TABL. The <size> is displayed in bytes.
For example, a sample of the response may look like:
560,8020,"Offset_1,TABL,220","Offset_2,TABL,340" ....
Naming Frequency Dependent Offset Tables
To rename a frequency dependent offset table use:
MEMory:TABLe:MOVE <string>,<string>
The first <string> parameter identifies the existing table name, and the
second identifies the new table name.
The following rules apply to frequency dependent offset table names:
a) Table names use a maximum of 12 characters.
b) All characters must be upper or lower case alphabetic characters,
or numeric (0-9), or an underscore (_).
c)
No spaces are allowed in the name.
Reviewing Table Data
To review the data stored in a frequency dependent offset table, use the
following commands:
• MEMory:TABLe:SELect "Offset1"
Select the sensor calibration table named "Offset1".
• MEMory:TABLe:SELect?
Query command which returns the name of the currently selected table.
• MEMory:TABLe:FREQuency:POINTs?
Query command which returns the number of stored frequency points.
• MEMory:TABLe:FREQuency?
Query command which returns the frequencies stored in the frequency
dependent offset table (in Hz).
• MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
Query command which returns the number of offset factor points stored
in the frequency dependent offset table.
EPM-P Series Power Meters Programming Guide