Sense[1]]|Sense2:Average:sdetect <Boolean> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:AVERage:SDETect <boolean>
This command enables and disables step detection. In AUTO filter mode,
the average of the last four values entered into the filter is compared to
the average of the entire filter. If the difference between the two averages
is greater than 12.5%, the digital filter is cleared. The filter then starts
storing new measurements. This feature is known as step detection and
shortens the settling time of the filter when the input power changes
substantially.
Note
Step detection is automatically disabled under the following
circumstances:
TRIG:DEL:AUTO is ON and the trigger mode is set to free run.
or,
Auto-averaging is disabled.
Under the above circumstances the value of SENS:AVER:SDET is
ignored. Note also that SENS:AVER:SDET is not set by the instrument
(that is, SENS:AVER:SDET will retain its current setting which may
indicate that step detection is ON).
Syntax
SENS
SENS2
Example
SENS:AVER:SDET OFF
Reset Condition
On reset, step detection is enabled.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:AVERage:SDETect <boolean>
:
:SDET
1
AVER
SENSe Subsystem
Space
0|OFF
1|ON
?
This command disables step
detection.
9-11