Agilent Technologies E8363BU-014 Installation Note
Agilent Technologies E8363BU-014 Installation Note

Agilent Technologies E8363BU-014 Installation Note

Configurable test set upgrade kit to add option 014

Advertisement

Quick Links

Installation Note
Configurable Test Set Upgrade Kit
To Add Option 014
Upgrade Kit Order Numbers: E8363BU-014, E8363CU-014,
E8364BU-014, and E8364CU-014
For E8364B/C PNA Series Microwave Network Analyzers
Agilent Kit Number: E8364-60103
Agilent Part Number: E8364-90019
Printed in USA June 2008
Supersedes print date: December 2007
© Agilent Technologies, Inc. 2003, 2004, 2007, 2008

Advertisement

loading

Summary of Contents for Agilent Technologies E8363BU-014

  • Page 1 Upgrade Kit Order Numbers: E8363BU-014, E8363CU-014, E8364BU-014, and E8364CU-014 For E8364B/C PNA Series Microwave Network Analyzers Agilent Kit Number: E8364-60103 Agilent Part Number: E8364-90019 Printed in USA June 2008 Supersedes print date: December 2007 © Agilent Technologies, Inc. 2003, 2004, 2007, 2008...
  • Page 2: Warranty Statement

    “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3: Contacting Agilent

    Contacting Agilent Assistance with test and measurements needs and information on finding a local Agilent office are available on the Web at: www.agilent.com/find/assist If you do not have access to the Internet, please contact your Agilent field engineer. In any correspondence or telephone conversation, refer to the Agilent product by NOTE its model number and full serial number.
  • Page 4 Description of Option 014 An Option 014 analyzer can be configured to measure high-power devices and devices that require high dynamic range. For a high-power measurement, external amplifiers and high power attenuators or isolators can be added to complete the test setup. In this configuration, test-port output-power up to 1 Watt (+30 dBm) can be applied to the device under test (DUT).
  • Page 5 Items Included in the Upgrade Kit Check the contents of your kit against this list. If any item is missing or damaged, contact Agilent Technologies. Refer to “Contacting Agilent” on page Table 1 Contents of Option 014 Upgrade Kit (E8364-60103)
  • Page 6 Table 1 Contents of Option 014 Upgrade Kit (E8364-60103) Description Part Number Desig. These parts are for analyzers WITH Option UNL AND Option 016 Lower front panel overlay (Option UNL/014/016) E8364-80024 PORT 1 RCVR A IN to A43 channel A step attenuator E8364-20129 PORT 2 RCVR B IN to A44 channel B step attenuator E8364-20130...
  • Page 7: Electrostatic Discharge Protection

    Installation Procedure for the Upgrade Kit The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation. This installation requires the removal of the analyzer’s protective outer WARNING covers.
  • Page 8: Overview Of The Installation Procedure

    Overview of the Installation Procedure Step 1. Remove the Outer Cover. Step 2. Remove the Front Panel Assembly. Step 3. Raise the Receiver Deck. Step 4. Remove the Existing Cables. Step 5. Install the Option 014 Cables. Step 6. Lower and Fasten the Receiver Deck. Step 7.
  • Page 9 Step 1. Remove the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. Refer to Figure 2 for this procedure.
  • Page 10 Step 2. Remove the Front Panel Assembly Refer to Figure 3 for this procedure. 1. With a T-10 TORX driver, remove the eight screws (item ) from the sides of the frame. Before removing the front panel from the analyzer, lift and support the front of CAUTION the analyzer chassis.
  • Page 11 Step 3. Raise the Receiver Deck Refer to Figure 4 for this procedure. 1. Place the analyzer bottom-side up on a flat surface. 2. With a T-10 TORX driver, remove the four screws, (item ), that secure the receiver deck. 3.
  • Page 12 Step 4. Remove the Existing Cables Analyzers WITHOUT Option UNL OR 016 Refer to Figure 5 for this procedure. Remove these cables in the order listed: • W7 E8364-20025 Channel R1 detector to A28 channel R1 mixer • W8 E8364-20026 Channel R2 detector to A29 channel R2 mixer •...
  • Page 13 Analyzers WITH Option UNL but WITHOUT Option 016 Refer to Figure 6 for the following procedure. Remove these cables in the order listed: • W8 E8364-20175 Channel R2 detector to A29 channel R2 mixer • W7 E8364-20174 Channel R1 detector to A28 channel R1 mixer •...
  • Page 14 Analyzers WITH Options UNL AND 016 Refer to Figure 7 for the following procedure. Remove these cables in the order listed: • W8 E8364-20150 Channel R2 detector to A29 channel R2 mixer • W7 E8364-20149 Channel R1 detector to A28 channel R1 mixer •...
  • Page 15 Step 5. Install the Option 014 Cables Analyzers WITHOUT Option UNL OR 016 Refer to Figure 8 for the following procedure. The new parts referenced in this procedure are listed in Table 1 on page Install these cables in the order listed. Use a 5/16-inch torque wrench set to 10 in-lbs. •...
  • Page 16 Analyzers WITH Option UNL but WITHOUT Option 016 Refer to Figure 9 for the following procedure. The new parts referenced in this procedure are listed in Table 1 on page Install these cables in the order listed. Use a 5/16-inch torque wrench set to 10 in-lbs. •...
  • Page 17 Analyzers WITH Options UNL AND 016 Refer to Figure 10 for the following procedure. The new parts referenced in this procedure are listed in Table 1 on page Install these cables in the order listed. Use a 5/16-inch torque wrench set to 10 in-lbs. •...
  • Page 18 Step 6. Lower and Fasten the Receiver Deck Refer to Figure 11 for this procedure. 1. Pull the latch pin (item ) toward the opposite side of the analyzer to release the receiver deck. 2. Lift the receiver deck to partially lower it, then release the latch pin (item ).
  • Page 19 Step 7. Replace the Lower Front Panel Overlay The new parts referenced in this procedure are listed in Table 1 on page NOTE Refer to Figure 12 for this procedure. 2. From the back side of the front panel, use a blunt object in one of the cutouts in the lower frame to push the overlay (item ) and separate it from the front panel.
  • Page 20 Step 8. Reinstall the Front Panel Assembly and Install Front Panel Jumpers Refer to Figure 13 for this procedure. 1. Lift and support the front of the analyzer chassis.. 2. Tighten all 12 of the front-panel feed-through connectors using a 5/16-inch torque wrench set to 21-in lbs. 3.
  • Page 21 Step 9. Reinstall the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. Refer to Figure 14 for this procedure.
  • Page 22 Step 10. Enable Option 014 Procedure Requirements • The analyzer must be powered up and operating to perform this procedure. • The Network Analyzer program must be running. • A mouse must be connected to the analyzer for this procedure. Option Enable Procedure 1.
  • Page 23 Step 11. Perform Post-Upgrade Adjustments and Calibration Adjustments The following adjustments must be made due to the hardware changes of the analyzer. • source calibration • receiver calibration These adjustments are described in the PNA service guide and in the PNA on-line HELP. A list of equipment required to perform these adjustments can be found at “Equipment Required for Post-Upgrade Adjustments”...
  • Page 24 Installation Note E8364-90019...