Sense[1]]|Sense2:Detector:function <Character_Data> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 sensor.
Syntax
SENS
SENS2
Parameters
Item
character_data
1. When measurement mode is set to average:
• If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
• INIT:CONT is set automatically to ON.
• SENS:AVER2:STAT is set automatically to OFF.
• CALC:FEED is set automatically to "POW:AVG" for all CALC blocks using
2. When measurement mode is set to NORMal:
• SENS:CORR:DCYC:STAT is set automatically to OFF.
Example
SENS1:DET:FUNC NORM
Reset Condition
On reset, the mode is set to NORMal.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
:
1
DET
Description/Default
Defines the measurement mode:
• AVERage: sets the E9320 sensor
to average only mode.
• NORMal: sets the E9320 sensor to
normal mode.
the specified channel in their CALC:MATH:EXPR.
:
Space
FUNC
character_data
Range of Values
AVERage
NORMal
This command sets the sensor to
normal mode for channel A.
SENSe Subsystem
?
1
2
9-41