Calibration Factor Block Layout
Calibration Factor Block Layout
E9300 Series Sensors:
No.
Calibration Factor
Bytes
Block Layout
1
Cal factor (high power)
2
These table entries are repeated as shown for each frequency point
Frequency (point 'N')
4
Cal factor (low power)
2
Cal factor (high power)
2
Table size:
-
Calibration Factor Block Layout: E9320 Series Sensors
E9320 Series Sensors:
No.
Calibration Factor
Bytes
Block Layout
Header:
Number of tables
1
Number of freq. points
2
Bytes per freq. point
1
Freq. LSB weight
2
Header Total:
Cal Factor Table:
Frequency (point '0')
4
A-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Contents
Data Format
Data Range
-
2.14
0.25 to 3
-
32 bit fixed
0 to Fhpb* (2^32)
-
2.14
0.25 to 3
-
2.14
0.25 to 3
See note
Contents
Data Format
Data Range
1
-
-
16 bit integer
-
-
1000
6
-
32 bit fixed
0 to Fhpb* (2^32)
Units
None
Power (in watts) is divided by
this value.
None
Fhbp = Freq Hz per bit.
None
Power (in watts) is divided by
this value.
None
Power (in watts) is divided by
this value.
The table size is dependent on
the number of freq. points.
Units
-
None
Number of cal factor tables.
This is currently unused but
has been set to a default value
of 1.
-
None
-
None
Number of bytes in cal factor
value at each freq.
-
Hertz
Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1KHz x 2^32 =
4.3E+12 = 4300 GHz range
None
Fhpb = Freq Hz per bit
Agilent E4416A/E4417A Programming Guide