Measurement Commands
FETCh[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
FETCh[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, recalculates the measurement and places
the result on the bus. The result is a power based measurement and is
expressed in the units defined by UNIT[1]|2|3|4:POWer.
Syntax
FETC
:SCAL
1
2
3
Space
4
Parameters
Refer to "Optional Parameters", on page 2-3 for additional details on the
parameters in this command.
Item
expected_value
(for the expected
power level)
resolution
2-24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:POW
:AC
?
,
expected_value
DEF
Description/Default
The expected power level
parameter can be set to DEF or a
numeric value. If a value is
entered it should correspond to
that set by CONFigure otherwise
an error occurs. The units of
measurement are dBm and W. The
default units are defined by
UNIT:POWer.
A numeric value for the resolution.
If it is unspecified the current
resolution setting is used. If a
value is entered it should
correspond to the current
resolution setting otherwise an
error occurs.
EPM-P Series Power Meters Programming Guide
,
resolution
source list
DEF
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF