Agilent Technologies EPM-P Series Programming Manual page 450

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SYSTem Subsystem
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
This command sets the GPIB address of the power meter.
Syntax
SYST
:COMM
Parameters
Item
numeric_value
Example
SYST:COMM:GPIB:ADDR 13
Query
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess? MIN|MAX
The query returns the current setting of the GPIB address or the values
associated with MIN and MAX.
11-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:GPIB
:SELF
:ADDR
Description/Default
A numeric value for the address.
the default value is 13.
DEF:
0.
MIN:
: 30.
MAX
Agilent E4416A/E4417A Programming Guide
Space
numeric_value
DEF
MIN
MAX
Space
?
MIN
MAX
Range of Values
0 to 30
DEF
MIN
MAX
This command sets the GPIB
address to 13.