Initiate[1]|2:Continuous <Boolean> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

INITiate[1]|2:CONTinuous <boolean>
This command sets the power meter for either a single trigger cycle or
continuous trigger cycles. A trigger cycle means that the power meter
exits the wait for trigger state and starts a measurement.
When entering local mode, if TRIGger[:SEQuence[1]|2]:SOURce is
set to INT[[1]|2] or EXT, INITiate:CONTinuous is not changed. For
other trigger sources, INITiate:CONTinuous is set to ON.
If INITiate:CONTinuous is set to:
Note
This command performs the same function as
INITiate:CONTinuous:SEQuence[1]|2 <boolean>.
Syntax
INIT
Example
INIT2:CONT ON
Reset Condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power-up, when entering
local mode, if TRIGger[:SEQuence[1]|2]:SOURce is set to
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OFF, the trigger system remains in the idle state until it is set to
ON, or INITiate:IMMediate is received. Once this trigger cycle is
complete the trigger system returns to the idle state.
ON, the trigger system is initiated and exits the idle state. On
completion of each trigger cycle, the trigger system immediately
commences another trigger cycle without entering the idle state.
:CONT
1
2
TRIGger Subsystem
INITiate[1]|2:CONTinuous <boolean>
Space
0|OFF
1|ON
?
This command places channel B in
the wait for trigger state.
13-5