[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time-gated period (time-gate length)
for time-gated measurements for E9320 sensors which are set to NORMal
mode. To set an E9320 sensor to NORMal mode, refer to the command
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
on page 9-41.
Syntax
SENS
SENS2
Parameters
Item
numeric_value
Example
SENS2:SWE3:TIME 0.001
Reset Condition
On reset, gate 1 is set to 100us and other gates to 1s.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME?
The query returns the current length of the time-gated period.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
:
1
SWE
1
2
3
4
Description/Default
The length of the time gated period in
seconds.
• DEF: the default value is 100 us
Units are resolved to 1 ns.
SENSe Subsystem
Space
: TIME
numeric_value
DEF
?
Range of Values
0 to 1 second
DEF
This command sets the length to
0.001 seconds.
9-55