Disp:wind1:Trace:low -35; Disp:wind1:Trace:upp 20; Sens:trac:offs:time -0.0002; Sens:trac:time - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Power Meter Remote Operation
Making Measurements on Wireless Communication Standards
TRIG:LEV:AUTO OFF
TRIG:LEV -20.00DBM
TRIG:HOLD 0.00065
TRIG:HYST 3.0

DISP:WIND1:TRACE:LOW -35

DISP:WIND1:TRACE:UPP 20

SENS:TRAC:OFFS:TIME -0.00001

SENS:TRAC:TIME 0.00065

DISP:WIND1:FORM TRACE

DISP:WIND2:FORM DNUM
CALC2:FEED1 "POW:AVER ON SWEEP1" Lower window upper
CALC4:FEED1 "POW:PEAK ON SWEEP1" Lower window lower
1-34
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Turn off auto leveling for
trigger
Sets trigger level to
-20.0 dBm
Sets trigger hold-off to
4650 μs
Sets Hysteresis to 3 dB
Sets trace display minimum
power to -35 dBm

Sets trace display maximum

power to +20 dBm

Trace starts 10 μs before
trigger point
Trace span set to 650 μs
Assigns upper window to a
trace display
Assigns lower window to a
dual numeric display
display line to show average
power using timing defined
by gate1
display line to show peak
power using timing defined
by gate1
EPM-P Series Power Meters Programming Guide