Agilent Technologies EPM-P Series Programming Manual page 276

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Parameters
Item
numeric_value
Example
DISP:WIND1:ANAL:LOW -50
Reset Condition
On reset, the value is set to -70 dBm for both windows.
Query
DISPlay:[WINDow[1]|2]:ANALog:LOW? [MIN|MAX]
The query returns the current setting of the analog meter's lower scale
limit, or the value associated with MIN or MAX. The format of the response
is <NR3>. The units in which the results are returned are determined by
the current setting of UNIT:POWer and CALCulate:RELative:STATe as
shown in Table 5-1.
Query Example
DISP:WIND1:ANAL:LOW?
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
DISPlay[:WINDow[1]|2]:ANALog:LOWer <numeric_value>
Description/Default
A numeric value for the analog meter
lower scale limit:
• DEF: the default is -70 dBm
• MIN: -150 dBm
• MAX: 230 dBm
Units used are determined by the
current setting of UNIT:POWer and
CALCulate:RELative:STATe as
shown in the previous table.
This command sets the upper
window's analog meter lower scale
limit to -50 dBm
This command queries the lower
scale limit set on the analog meter
in the upper window.
DISPlay Subsystem
Range of Values
to
dBm
-150
230
DEF
MIN
MAX
5-11