Agilent Technologies EPM-P Series Programming Manual page 600

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

IEEE488.2 Command Reference
*DDT <arbitrary block program data>|<string program data>
*DDT <arbitrary block program data>|<string program data>
The *DDT (Define Device Trigger) command determines the power meter's
response to a GET (Group Execute Trigger) message or *TRG common
command. This command effectively turns GET and *TRG into queries,
with the measured power being returned.
Syntax
*DDT
Parameters
Type
arbitrary block program
data
string program data
1. The <action> field of the parameter may contain:
2. The first digit after the # indicates the number of following digits. The following
digits indicate the length of the data.
Examples of <arbitrary block program data> parameters are:
• #15FETC? and #206FETCh?
Examples of <string program data> are:
• "FETCh1?", "FETCh?" and "TRIG1;FETC1"
Note
*DDT does not support "FETCh1?" and "FETCh2?"
simultaneously.
16-8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
arbitrary block program data
string program data
?
Description
The command which is
executed on a GET or
*TRG.
• FETC?
• FETC1?
• FETC2? (E4417A only)
*TRG
TRIG1
(E4417A only)
TRIG2
EPM-P Series Power Meters Programming Guide
Range of Values
1,2
#nN<action>
1
"<action>"

Advertisement

Table of Contents
loading

Table of Contents