SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor
data from, the currently connected sensor. The whole calibration factor
block must be written at once as a checksum is generated. The new block
must not be larger than the existing block.
This command applies to E9320 Series sensors for peak path data only.
For E4410 Series, E9300 Series, and E9320 Series sensors, average path
data, refer to SERVice:SENSor[1]|2:CALFactor
<cal_factor_data>.
Syntax
SERV
Parameters
Item
cal_factor_data
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
Query Example
SERV:SENS:PCALf?
Error Messages
• If no power sensor is connected, error -241 "Hardware missing" occurs.
• If a non E9320 Series sensor is connected, error -241
"Hardware missing" occurs.
• If INIT:CONT is not set to OFF, error -221, "Settings conflict" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
:SENS
1
:PCALf
2
Description/Default
A binary data block. Refer to
Appendix A for further information.
SERVice Subsystem
Space
cal_factor_data
?
This command returns the peak
path calibration factor block for
channel A.
15-19