Parameters
Item
numeric_value
Example
AVER:COUN 400
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values
associated with MIN and MAX. The format of the response is <NR1>.
Query Example
AVER:COUN?
Error Messages
If a filter length value is entered using
[SENSe[1]]|SENSe2:AVERage:COUNt while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings
Conflict" occurs. However, the filter length value is set but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically
set ON.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
Description/Default
A numeric value defining the filter length.
• DEF: the default value is 4.
• MIN: 1.
• MAX: 1024.
This command enters a filter
length of 400 for channel A.
This command queries the filter
length for channel A.
SENSe Subsystem
Range of Values
1 to 1024
DEF
MIN
MAX