Power Meter Remote Operation
Using Frequency Dependent Offset Tables
• MEMory:TABLe:GAIN[:MAGNitude]?
Query command which returns the offset factors stored in the frequency
dependent offset table.
Modifying Data
If you need to modify the frequency and offset factor data stored in a
frequency dependent offset table you need to resend the complete data lists.
There are two ways to do this:
1. If you have retained the original data in a program, edit the
program and resend the data.
2. Use the query commands shown in "Reviewing Table Data", on
page 1-48 to enter the data into your computer. Edit this data, then
resend it.

Selecting a Frequency Dependent Offset Table

After you have created the frequency dependent offset table, you can select
it using the following command:
[SENSe[1]]|SENSe2:CORRection:CSET2[:SELect] <string>
To find out which frequency dependent offset table is currently selected, use
the query:
[SENSe[1]]|SENSe2:CORRection:CSET2[:SELect]?
Enabling A Frequency Dependent Offset Table
To enable the frequency dependent offset table, use the following command:
[SENSe[1]]|SENSe2:CORRection:CSET2:STATe ON
If you set [SENSe[1]]|SENSe2:CORRection:CSET2:STATe to ON and no
frequency dependent offset table is selected error -221, "Settings conflict"
occurs.
Making The Measurement
To make the power measurement, set the power meter for the frequency of
the signal you want to measure. The power meter automatically sets the
1-56