Syntax
SENS
1
SENS2
Parameters
Item
numeric_value
Example
CORR:DCYC 90PCT
Reset Condition
On reset, the duty cycle is set to 1% (DEF).
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]? [MIN|MAX]
The query returns the current setting of the duty cycle or the values
associated with MIN and MAX.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]
:
CORR
:DCYC
:GAIN3
Description/Default
A numeric value for the duty cycle.
• DEF: the default value is 1%.
• MIN: 0.001%.
• MAX: 99.999%.
The units are PCT, and are optional.
SENSe Subsystem
:INP
:MAGN
Space
numeric_value
DEF
MIN
MAX
Space
?
MIN
MAX
Range of Values
0.001 to 99.999
DEF
MIN
MAX
This command sets a duty cycle of
90% for channel A.
PCT
9-31