MEMory:CATalog[:ALL]?

This command lists stored sensor calibration tables (8480 Series power
sensors and N8480 Series sensors with Option CFT only), frequency
dependent offset tables and save/recall registers.
The power meter returns the data in the form of two numeric parameters
and as many strings as there are stored tables and save/recall registers:
<numeric_value>,<numeric_value>{,<string>}
A sample of a response may look like the following:
1178,26230,"DEFAULT,TABL,14","HP8481A,TABL,116",
"HP8482A,TABL,74",..........."State0,STAT,1619",
"State1,STAT,1619","State2,STAT,1619" ...........
Note
Predefined sensor calibration table is not applicable for N8480 Series
power sensors with Option CFT. Therefore you are required to create a
new sensor calibration table for the sensors when a sensor calibration
table is needed.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The first numeric parameter indicates the amount of memory, in
bytes, used for the storage of tables and registers.
The second numeric parameter indicates the memory, in bytes,
available for the storage of tables and registers.
Each string parameter returned indicates the name, type and size
of a stored table or save/recall register:
<string>,<type>,<size>
<string> indicates the name of the table or save/recall
register.
<type> indicates TABL for sensor calibration and frequency
dependent offset tables, or STAT for a save/recall register.
<size> indicates the size of the table or save/recall register
in bytes.
MEMory Subsystem
MEMory:CATalog[:ALL]?
7-5