OUTPut:RECorder[1]|2:STATe <boolean>
This command enables or disables the specified recorder output.
Syntax
OUTP
Example
OUTP:REC1:STAT 1
Reset Condition
On reset, the recorder output is OFF.
Query
OUTPut:RECorder[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the specified recorder is switched on.
Query Example
OUTP:REC2:STAT?
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
:REC
2
1 is returned when the recorder output is switched ON.
0 is returned when the recorder output is switched OFF.
OUTput Subsystem