TRIGger Subsystem
TRIGger[:SEQuence]:HOLDoff <numeric_value>
TRIGger[:SEQuence]:HOLDoff <numeric_value>
This command sets the trigger holdoff in seconds.
Note
The command is accepted for TRIGger[:SEQuence[1]] (channel A)
only, for both single and dual channel power meters.
Syntax
TRIG
Parameters
Item
numeric_value
Example
TRIG:SEQ1:HOLD 0.1
Reset Condition
On reset the trigger holdoff is set to 1 us.
13-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:HOLD
Space
:SEQ
Description/Default
The trigger holdoff in seconds.
DEF: the
default value is 1 us.
• MIN: 1 us.
• MAX: 400 ms.
Units are resolved to 100 ns.
This command sets the trigger
holdoff to 100 ms for channel A.
EPM-P Series Power Meters Programming Guide
numeric_value
DEF
MIN
MAX
?
Range of Values
1 us to 0.4 seconds
DEF
MIN
MAX