TRIGger Subsystem
TRIGger[:SEQuence]:LEVel <numeric_value>
TRIGger[:SEQuence]:LEVel <numeric_value>
This command sets the power level at which a trigger event is recognized.
Note
The command is accepted for TRIGger[:SEQuence[1]] (channel A)
only, for both single and dual channel power meters.
Syntax
TRIG
Example
TRIG:SEQ:LEV 10
Parameters
Item
numeric_value
Reset Condition
On reset the power level is set to 0 dBm.
Query
TRIGger[:SEQuence]:LEVel?
The query returns the current power level setting.
13-26
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:LEV
Space
:SEQ
This command sets the power level
for a trigger event to 10 dBm
Description/Default
The power level at which a trigger
event is recognized.
the default value is 0 dBm.
DEF:
Units are resolved to 0.1 dBm.
EPM-P Series Power Meters Programming Guide
numeric_value
DEF
?
Range of Values
-40 to 20 dBm
DEF