Power Meter Remote Operation
Making Measurements on Wireless Communication Standards
TRIG:LEV:AUTO OFF
TRIG:LEV -10.00DBM
DISP:WIND1:FORM SNUM
DISP:WIND2:FORM DNUM
CALC1:FEED1 "POW:AVER"
CALC2:FEED1 "POW:PEAK"
CALC4:FEED1 "POW:PTAV"
1-40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Turn off auto leveling for
trigger
Sets trigger level to
-10.0 dBm
Assigns upper window to a
single numeric display
Assigns lower window to a
dual numeric display
Upper window to show
average power
Lower window upper
display line to show peak
power
Lower window lower
display line to show
peak-to-average ratio
EPM-P Series Power Meters Programming Guide