Agilent Technologies EPM-P Series Programming Manual page 165

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Measurement Commands
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence:RELative
[<expected_value>[,<resolution>[,<source list>]]]
Item
source list
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right. The
parameter DEFault is used as a place holder. Specifying
value unchanged.
2. When the measurement result is linear this parameter represents the number of
significant digits. When the measurement result is logarithmic 1 to 4 represents a
resolution of 1, 0.1, 0.01 and 0.001 respectively.
3. E4417A only.
Example
CONF1:DIFF:REL DEF,1,(@1),(@2)
2-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
This channel list specifies the
channels used to calculate the
difference.
If unspecified and the current
window setup is a difference
measurement then this difference
setup is used, otherwise it defaults
to channel A-B (E4417A) or A-A
(E4416A).
DEF
means DEFault. This is not equivalent to the DEFault
This command configures the
upper window to make a
difference measurement of
channel A - channel B with
relative mode on, using the
current sensor range and a
resolution of 1 on both channels.
EPM-P Series Power Meters Programming Guide
Range of Values
3
(@1),(@2)
3
(@2),(@1)
(@1),(@1)
3
(@2),(@2)
DEF
leaves the parameter