CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA] <numeric_value>
Parameters
Item
numeric_value
Example
CALC2:LIM:LOW:DATA 0.1
Reset Condition
On reset, both measurements of the upper and lower windows are set to
-90.00 dBm or -90 dB (DEF).
Query
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA]? [MIN|MAX]
The query returns the current setting of the lower limit or the values
associated with MIN and MAX for the specified window.
Query Example
CALC2:LIM:LOW:DATA?
3-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the lower test
limit:
• DEF: the default is -90.00 dBm
or -90 db.
• MIN: -150 dBm or -180 dB.
• MAX: +230 dBm or +200 dB.
This command enters a lower limit
for the lower window/upper
measurement depending on the
window's units as follows:
dBm = 0.1 dBm
W = 100 mW
dB = 0.1 dB
% = 0.1 %
This command queries the lower
limit set for the lower window
upper measurement.
EPM-P Series Power Meters Programming Guide
Range of Values
-150 to +230 dBm or
-180 to +200 dB
DEF
MIN
MAX