System:communicate:serial:control:dtr <Boolean> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SYSTem:COMMunicate:SERial:CONTrol:DTR <boolean>
This command sets the serial interface (RS232) DTR signal line either
high always (ON) or low always (OFF). In addition, the DTR signal line can
be tied to the condition of the interface receiver buffer (IBFull) resulting
in the line going high when the receiver is ready to accept data and low
when the receiver buffer is full (not ready for data). Also, with DTR in
IBFull mode, the transmitter will be inhibited when DSR is low.
Syntax
SYST
Example
SYST:COMM:SER:CONT:DTR ON
Reset Condition
On reset, the DTR signal line setting is unaffected.
Query
SYSTem:COMMunicate:SERial:CONTrol:DTR?
The query returns the current DTR signal line setting, ON, OFF or IBFull.
Query Example
SYST:COMM:SER:CONT:DTR?
Error messages
If the RS422 interface is selected, then the error message -221 "Settings
conflict" will occur.
Agilent E4416A/E4417A Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:COMMunicate:SERial:CONTrol:DTR <boolean>
:COMM
:SER
:CONT
SYSTem Subsystem
Space
:DTR
ON
OFF
IBFull
?
This command sets the DTR
signal line high (always).
This command queries the setting
of the DTR signal line.
11-7