READ[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence? [<expected_value>[,<resolution>[,<source list>]]]
READ[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to
difference mode with relative mode off, aborts then initiates both channel
A and B, calculates the difference measurement result and places the
result on the bus. The result is a power based measurement and is
expressed in the units defined by UNIT[1]|2|3|4:POWer.
INITiate:CONTinuous must be set to OFF on both channels,
Note
otherwise error -213, "INIT ignored" occurs. If TRIGger:SOURce
is set to BUS on either channel, error -214, "Trigger deadlock"
occurs.
Syntax
READ
1
:SCAL
2
3
Space
4
Parameters
Refer to "Optional Parameters", on page 2-3 for additional details on the
parameters in this command.
Item

expected_value

(for the expected
power level)
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:POW
:AC
:DIFF
,
expected_value
DEF
Description/Default
The power meter ignores the
numeric value entered in this
parameter. Any value entered is
treated like DEF.
Measurement Commands
?
,
resolution
source list
DEF
Range of Values
sensor dependent
1
DEF