MEMory Subsystem

MEMory:CATalog:TABLe?

MEMory:CATalog:TABLe?
This command is used to list the stored sensor calibration (8480 Series
power sensors and N8480 Series sensors with Option CFT only) and
frequency dependent offset tables.
The power meter returns the data in the form of two numeric parameters
and as many strings as there are stored tables.
<numeric_value>,<numeric_value>{,<string>}
For example, a sample of a response may look like:
1178,10040,"DEFAULT,TABL,14","HP8481A,TABL,116",
"HP8482A,TABL,74","HP8483A,TABL,62"...........
Note
Predefined sensor calibration table is not applicable for N8480 Series
power sensors with Option CFT. Therefore you are required to create a
new sensor calibration table for the sensors when a sensor calibration
table is needed.
7-8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The first numeric parameter indicates the amount of memory, in
bytes, used for the storage of tables.
The second parameter indicates the memory, in bytes, available
for the storage of tables.
Each string parameter returned indicates the name, type and size
of a stored table:
<string>,<type>,<size>
<string> indicates the name of the table.
<type> indicates TABL for a table.
<size> indicates the size of the table in bytes.
EPM-P Series Power Meters Programming Guide