[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <boolean>
This command is used to enable and disable the use of the currently active
sensor calibration table (CSET1) or frequency dependent offset table
(CSET2). When a table has been selected and enabled, the calibration
factors/offsets stored in it can be used by specifying the required frequency
using the [SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is
taken from the sensor calibration table and is used during calibration.
The CSET1 command is only applicable for 8480 Series power sensors and
N8480 Series power sensors with Option CFT.
Syntax
:
SENS
1
SENS2
Example
CORR:CSET1:STAT 1
Reset Condition
On reset, the sensor calibration table and frequency dependent offset table
are not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a
table is enabled or disabled.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <boolean>
CORR
1
:CSET
:CSET2
1 is returned when the table is enabled.
0 is returned when the table is disabled.
SENSe Subsystem
Space
:STAT
0|OFF
1|ON
?
This command enables the use of
the currently active sensor
calibration table for channel A.
9-27