Display[:Window[1]|2][:Numeric[1]|2]:Resolution Numeric_Value> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

DISPlay[:WINDow[1]|2][:NUMeric[1]|2]:RESolution <numeric_value>
This command sets the resolution of the measurement result in the
specified window.
Syntax
DISP
:WIND
1
2
Parameters
Item
numeric_value
Example
DISP:WIND2:RES 4
Reset Condition
On reset, the resolution is set to 3.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
DISPlay[:WINDow[1]|2][:NUMeric[1]|2]:RESolution <numeric_value>
:NUM
1
:RES
2
Description/Default
A numeric value for the window
resolution:
• DEF: 3
• MIN: 1
• MAX: 4
DISPlay Subsystem
Space
numeric_value
DEF
MIN
MAX
Space
?
MIN
MAX
Range of Values
1 to 4
DEF
MIN
MAX
This command sets the lower
window's resolution to four
significant digits if the
measurement result is linear, or to
0.001 if the measurement result is
logarithmic.
5-21