Advertisement

SENSe Subsystem

SENSe[1]]|SENSe2:SWEep[1]|2|3|4 Commands

SENSe[1]]|SENSe2:SWEep[1]|2|3|4 Commands
These commands set offset time and time gate length as illustrated in the
following diagram:
Incoming signal
from sensor
Offset time and time gate length values can be set for up to four
measurement gates per channel. Measurement gate number is defined by
the numeric value following the SWEep component of the command.
These commands can only be used with E9320 sensors which must be
Note
set to NORMal mode.
The following commands are detailed in this section:
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME
<numeric_value>
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
9-52
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Pre-
Trigger
With no trace: internally calculated time
With no
trace:
internally
calculated
time
Trigger
Delay
TRIG:DEL
Delayed
Trigger
Point
Trigger
SENSe:SWEep:OFFSet:TIME
Point
Defined using:
TRIG:LEVel
TRIG:SLOPe
TRIG:HYSTeresis
Data Collection Time
EPM-P Series Power Meters Programming Guide
Post-Trigger
Time gate length:
SENSe:SWEep:TIME
Offset time: