OUTput Subsystem
OUTPut:TTL[1|2]:STATe <boolean>
OUTPut:TTL[1|2]:STATe <boolean>
This command is used to enable and disable the rear panel TTL Outputs.
There are two TTL outputs:
Both can be connected to any of the CALC subsytems (that is the upper
window upper/lower measurement or the lower window upper/lower
measurement) using OUTput:TTL:FEED <string>.
When enabled, the TTL Outputs can be used to determine when a
predefined limit is exceeded.
Syntax
Example
OUTP:TTL1:STAT 1
Reset Condition
On reset, the TTL outputs are disabled.
Query
OUTPut:TTL[1|2]:STATe?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the TTL output is enabled.
8-16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Output 1 (TTL1).
Output 2 (TTL2).
OUTP
:TTL
1
2
1 is returned when the TTL output is enabled.
0 is returned when the TTL output is disabled.
EPM-P Series Power Meters Programming Guide
Space
0|OFF
:STAT
1|ON
?
This command enables the rear
panel upper window TTL output.