Status:operation:sense[:Summary] - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

STATus:OPERation:SENSe[:SUMMary]

The operation status sense summary register set contains information on
the status of the power sensors.
The following bits in these registers are used by the power meter:
Bit
Number
0
1
2
3-15
These bits are set when the power meter is reading data from the
Agilent E-Series or N8480 Series power sensor EEPROM.
These bits are cleared when the power meter is not reading data from the
Agilent E-Series or N8480 Series power sensor EEPROM.
Syntax
STAT
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decimal
Definition
Weight
-
Not used
2
Channel A SENSe Status
4
Channel B SENSe Status (E4417A only)
-
Not used
:SENS
:SUMM
:OPER
STATus Subsystem
STATus:OPERation:SENSe[:SUMMary]
10-15